3838NCCT2.0 hW 80 L 40ASPECTS 10RL80004.524 hNO BLOODLYTIC: GOASPECTS 10looks normal
Fan beam: tube to detector arc through the head; it cuts the slice the monitor showsRotor: X-ray tube, collimator, 11 detector modules, DAS boards, generator, coolerContrast line: injector syringes (contrast, saline) to the Y, up over the table edge to the IV in the right armA stroke-protocol CT with the gantry cover off. Non-contrast CT first: blood is bright (50 to 80 HU) from the first minute, so a bleed rules out a clot-buster at once. Ischemic brain hides at first; it darkens only as the core takes up water, which is why the slice can look normal at 2 h and obvious at 12. The injector then pushes contrast for CTA, which finds the blocked artery, and for perfusion, which weighs the dead core against the brain still worth saving. The timer is hours since last known well; the readout is what the team does next.
Patient
Scan
2.0 h
3838NCCT2.0 hW 80 L 40ASPECTS 10RL
Axial slice at the basal ganglia. Viewer's right is the patient's left.

Lytic eligible

2.0 h after onset. No blood; no visible infarct yet (the core is only 1.8 HU darker, under what a reader sees in this window); hyperdense left MCA from the clot itself. No blood, inside 4.5 h: give IV alteplase or tenecteplase now, then CTA for a large-vessel occlusion.

Gray matter
38 HU
Core gray matter
--
Water uptake
--
ASPECTS read / true
--
Core · at risk
--
Blood
--

Why CT first

Stroke is either a blocked artery (ischemic, about 85 %) or a burst one (hemorrhage, about 15 %), and the two look the same at the bedside: sudden weakness, numbness, speech or vision loss. The treatments are opposite. A clot-buster that opens a blocked artery makes a bleed worse. So the first scan does not need to show the stroke; it needs to show whether there is blood. Non-contrast CT does that in under a minute with near-perfect sensitivity for an acute bleed, which is why the stroke pathway runs straight from the door to the scanner (target: imaging within 20 minutes of arrival, needle within 60, ideally 45). Only a finger-stick glucose is needed before it, because low sugar mimics stroke.

How the picture is made

The tube on the rotor throws a thin fan of X-rays through the head to the detector arc opposite. The rotor turns about twice a second; each detector channel records how much of the beam got through at each angle, and the computer back-projects a thousand such views into a map of attenuation in Hounsfield units: water 0, air −1000, bone over +1000. Brain sits in a narrow band: CSF about 5, white matter about 28, gray matter about 38, fresh blood 50 to 80. A window maps a slice of that range to black through white. The brain window (width 80, center 40) spreads 0 to 80 HU across the gray scale; the narrower stroke window (width 40, center 35) doubles the contrast, so a few HU of early swelling becomes visible.

What each scan answers

Time windows

Keyboard

Focus the figure. Click it or press 1, 2, 3 for NCCT, CTA, CTP. ← → change hours since onset (Shift for 2 h), B switches bleed and clot, W switches window, Space stops the tour.

Teaching model, not for clinical use. Water uptake 30 % × (1 − e^(−t/12 h)); core 10 + 110 × (1 − e^(−t/10 h)) mL; hematoma 14 to 24 mL, 56 to 78 HU. A reader is assumed to see a patch once its contrast reaches 4 % of the window.