Normal sinus rhythm.
One beat, taken apart
Every wave is one stage of a single impulse that starts in the SA node and ends with the ventricles recharged. The moving line is where the phantom is in the beat right now.
Six limb leads, one heart
A lead is a direction. Each one reads the shadow of the same heart vector on its own axis: impulse heading toward the lead's positive pole, pen up; heading away, pen down; square to it, a small up-and-down. That is why one beat looks different in every lead, and why the leads can find the axis.
Einthoven's law holds exactly: II = I + III, and aVR + aVL + aVF = 0. The augmented leads compare one limb with the average of the other two and read √3⁄2 of the projection. The shaded sector is the normal axis; the grey arrow is this rhythm's mean QRS axis; the bold line is the heart vector right now.
Measure it
Reading any ECG
The paper
At 25 mm/s and 10 mm/mV, one small box is 1 mm: 0.04 s across and 0.1 mV up. One big box is 5 mm: 0.2 s and 0.5 mV. Five big boxes are one second. The square step at the start of the strip is the 1 mV calibration: it should be two big boxes tall. Change the gain or speed above and watch the same beat grow or stretch.
In order, every time
- Rate: 300 ÷ big boxes between two R waves (1500 ÷ small boxes). Irregular: count QRS in a 6 s strip × 10.
- Rhythm: a P before every QRS, a QRS after every P, the same PR each time, P upright in II: sinus.
- Axis: QRS up in I and up in aVF is normal. Up in I, down in aVF: check II; down in II too is left axis deviation.
- Intervals: PR 120 to 200 ms, QRS under 120 ms, QTc under about 450 ms (men) or 470 ms (women).
- Waves: P shape; pathological Q waves (wider than 40 ms or deeper than a quarter of the R); QRS voltage; R-wave growth V1 to V6; ST shifts; T inversions.
- Compare with the last ECG. New is what matters.
Where each lead looks
| Leads | Wall | Usual artery |
|---|---|---|
| II, III, aVF | Inferior | Right coronary (most), circumflex |
| I, aVL, V5, V6 | Lateral | Circumflex, diagonal branches of the LAD |
| V1, V2 | Septal | LAD (septal branches) |
| V3, V4 | Anterior | LAD |
| aVR | Looks into the cavity | Up with ST down everywhere else: left main or three-vessel ischaemia |
The other six: chest leads
V1 to V6 are single electrodes across the chest, each read against Wilson's central terminal, the average of RA, LA and LL. They work exactly like the phantom, but in the horizontal plane: V1 sits over the right ventricle and sees the vector heading away (mostly down), V6 sits over the left ventricle and sees it coming (mostly up), and the R wave grows from V1 to V6. Limb leads plus chest leads are the 12-lead ECG: the same vector seen from twelve directions in two planes.
What the model is
A teaching model, not for clinical use. The heart is one dipole in the frontal plane, a sum of Gaussian lobes for P, the septal, free-wall and basal parts of QRS, an ST injury current and an asymmetric T; the electrodes read Φ = (v · r) ⁄ √3 at the corners of an equilateral Einthoven triangle, and the recorder only subtracts them. QT follows Bazett. Rhythm presets change the lobes, the timing and which conduction paths carry the impulse.